SuSI2 Filter Setup |
| Author: |
Gabriel Martin |
| Approved: |
O. Hainaut |
| Version: |
20 Nov. 1998 |
The filters used on SUSI2 are circular with a diameter of 100 mm.
It is also possible to use circular filter of diameter 85mm , in which case
a special adaptor cell has to be used , with variable flux loss in the field
and vignetting. Check out the special rules at the end of this file.
1. Special rules for non-standard filters
- Rule 1- do NOT mount EMMI (or any)
filters in SuSI unless approved by OPC or if there is a compelling reason
(e.g funny object that has to be observe absolutely at high resolution in
a special filter), approved by Team Leader or astronomer.
- Rule 2- if filters have to be mounted in the SuSI's filter
holders, only Gabriel Martin and the optics
team can do it.
2. Installation Procedure
- Press Rotator A emergency stop button at the right side of platform A .
- Get the boxes containing the filters you want to mount and the magnet
tool from SUSI2 filter red cabinet.
- Watching at co-rotator rotate carefully Rotator A until integrating
(yellow) sphere is on the right side .
- Unscrew the knob on left side of SUSI2 horizontal cover and open it.
- Start SUSI OS from INS start mouse button.
- If this fails startup fcd.
- Start SUSI config panel from INS submenu.
- On the SUSI2 configuration panel, select filter pos and click move
to load .
- Stick the magnet on the side of the filter housing ring and gently pull
until you can seize the ring by hand : store it in its box and screw in the
lid.
- Mount the new one with label outside . If a filter is dusty clean it with
some pressurized CO2.
- Store removed filter into red cabinet.
- Release the emergency stop button.
- Declare the new configuration in the SUSI2 configurarion panel.