RILD Focussing
RILD Focus must be check at every setup.
It need pinhole to be mounted.
- If not::
- Press EMMI emergency stop button at the center of EMMI cover.
- Remove the red side cover of EMMI :
- From the EMMI configuration panel, select a free starplate position
and click "move to load".
- Install pinhole in position.
- From EMMI configuration panel, select click enter starplate .
- Click Backup under File submenu.
- Release the emergency stop button.
- Run RILD_img_tec_SetFoc.
- When finishing it will pop up a menu pproposing to change or not the
focus formula constant term ( focr best focus encoder position at 0 degrees
C )
Only accept it if the focus curve in midas graphics window looks OK and the
difference between old and new values is less than 100 and more than 10.
[ last reference around 620 e.u.]
- If this difference is steadily more than hundred , report to EMMI instrunment
nucleus.