Procedure: |
SofI Wheel Alignment: normal procedure
|
| Keywords: |
Recovery - SofI - slit wheel - hardware init - fallBack init |
| Prepared/Revised: |
2004-06-08 ohainaut |
Reviewed: |
2004-06-09 uweilenm |
Released: |
2004-06-09 ohainaut |
| |
|
Revision History:
2004-06-07 ohainaut revision by Ueli and release
2004-06-06 ohainaut creation
|
Summary
This procedure describes who to re-align one of the SofI filter wheels that
has lost its init. This is the standard procedure, to be used when the
hardware init works. When hardware init works, the "Fallback init"
should NOT be used. See at the end how to recover if you "fallback
init" while you should not.
Procedure
1. Alignment of a wheel
- If needed, e.g. you are lost, perform a hardware init:
- click on the title of the wheel on the OS to open the motor panel
- if needed, select "backup init system" if the primary init
switch does not work (as of 2004-06-08: all wheels use primary, except
Slit/Mask which uses backup).
- [Hardware Init]
- The wheel should move to search its init point. If it does not move,
- if you have done a "fall back init" (you should not),
see procedure at the end of this page to reset the init.
- if not, you are in trouble; call instrumentation engineer.
- Align the wheel or the optical element. The method is different for each
wheel:
- For slit: select the slit, run the aligh-slit template
- For filter: Identify filter and center it
- For objective: identify and align together with a known slit
- For grism: identify and align grism, run spec align template
- Write down the encoder unit value corresponding to the aligned position.
- Menu Tool > ICS Config.
- In ICS config, [edit] the instrument file; set the encoder value corresponding
to the element you aligned to the value you obtained earlier. Save and exit.
- In ICS config, press [LOAD] to load the new value
- In motor panel, press [OFF] [STANDBY] [ONLINE] to load the new value and
to init the function at the correct position.
- check by doing a set-up to the slit or element you just aligned and check
that it is really aligned.
2. Reset after a (wrong) Fall-Back Init
In case you did a Fall-Back Init (while you should not have done so),
reset the system:
- Open the motor panel (click title of the wheel)
- Open FallBack Init panel: [Fall Back Init]
- Click [RESTORE]
This will establish the normal Init procedure and remove the Fall Back Init.