Procedure: |
SofI Filter Wheel Alignment at NIGHT TIME
|
| Keywords: |
Recovery - SofI - Filter wheel - fallBack init |
| Prepared/Revised: |
2002-06-22 vdoublie |
Reviewed: |
2002-06-22 lvanzi |
Released: |
lvanzi 2002-06-22 |
| |
|
Revision History:
2004-06-06 ohainaut reformat and comments
2002-06-22 vdoublie creation
|
Summary
This procedure describes who to re-align quickly one of the SofI filter wheel
during night time, in case the HARDWARE INIT does not work.
Note that this procedure should NOT be used in case the hardware init
can be used. It should be used ONLY when the filter wheel is in "FallBack
Init" condition, i.e. when the hardware init switches are failing.
Using this procedure when the hardware init should be used will screw up the
system.
Procedure
- A. The filter wheel is ONLINE (if not jump
to B)
- from OS: select Imaging Mode, Large Field, and
Ks filter
- Put the guide probe in Center Field
- go to a bright star (Bright star catalogue) of magnitude 3 < V <
5: center moving the telescope
- from the OS: offset telescope to send the star off the field of view
of SOFI using combined offsets (e.g. x ~ 600pix, y~ 600pix)
- Guide probe to Park
- Store a Fixed Pattern, and turn it ON
- Stop AO, then defocus telescope (M2) by 14 encoder steps: check on
guide camera that the star looks like a "doughnut"
- offset BACK to the star from the OS using offset ( e.g. x = -600, y
= -600): You should see the image of the telescope pupil
- if the filter is misaligned you should see something like display in
figure #2 below
- if the filter wheel does not seem to vignet the pupil image,
- a) The filter wheel may not be misaligned, to check refer to DayTime
recovery procedure A.2. If the spectrum does not correspond, then
- b) on the OS panel change filter one by one and carefully monitor
the count on the pupil image: Ks has the highest counts, while the
spectroscopic filters GBF and GRF will saturate. Then follow Day
Time recovery procedure A.2
- If needed, Go back on sky (Calibration mirror OUT), using the ICS Motor
Device panel, align such as: find the encoder limits for left-side vignetting,
and for the right-side vignetting, then position the wheel at the intermediate
encoder value
- then do a Fall Back Initialisation
at the Ks position
- B. The filter wheel is NOT ONLINE
from the ICS Motor Device panel, put the
wheel OFF, then STANDBY. The recovery follows mainly the same steps
as if the wheel were ONLINE, but moving the wheel is done using the Handset
Tool. Because the whell is NOT ONLINE, we will assumed that the filter position
has been lost.
- from OS: select Imaging Mode, Large Field, and
Ks filter
- Put the guide probe in Center Field
- go to a bright star (Bright star catalogue) of magnitude 3 < V <
5: center moving the telescope
- from the OS: offset telescope to send the star off the field of view
of SOFI using combined offsets (e.g. x ~ 600pix, y~ 600pix)
- Guide probe to Park
- Store a Fixed Pattern, and turn it ON
- Stop AO, then defocus telescope (M2) by 14 encoder steps: check on
guide camera that the star looks like a "doughnut"
- offset BACK to the star from the OS using offset ( e.g. x = -600, y
= -600):
- Use the handset tool to move the filter wheel. Again Ks should show
the highest counts level, while the filters GRF and GBF should saturate
(don't stay at these positions too long to avoid damaging the detector).
- Align such as: find the encoder limits for left-side vignetting, and
for the right-side vignetting, then position the wheel at the intermediate
encoder value
- Once you think you are on the Ks filter, perform the Fall
Back Initialisation