| Procedure: | SOFI Filter Wheel Align |
| Author: | Leo Vanzi & V. Doublier |
| Release: | 1.0/ June 22, 2002 |
This procedure apply if the filter wheel becomes misaligned during the day time activities (Cal Plan, restart, of ICS ...):
The
filter wheel does NOT initialise!
Make sure that the Nasmyth shutter in CLOSED and Calibration Unit is IN
A- The filter wheel is still ONLINE (if not jump to B)
1- Align roughly the filter wheel:
- from OS: select Imaging Mode, Large Field, and Ks filter
- open ICS motor device control panel, select filter 1 function
- On RTD open a statistics box, and move it to a relatively hot pixel free area and close to one edge.
- look at the count: in the Ks filter the average counts should be around 2000 in morning, and 3000 in the afternoon: Ks is very sensitive to the thermal background.
- if the counts are below 1000, or if the detector is saturating: the filter wheel is badly mis-aligned
- using the Move Enc. from the ICS Motor Device panel. Move only by few encoder offsets between positions.
- Once you found the Ks filter, reduce the encoder movement and maximize the count level within the statistics box. Because the statistics box is close to the edge it will be sensitive to the "vigneting" due to the edge of the filter holder.
- Do a Fall Back Initialisation at the Ks position
2- Fine Tuning:
- From the OS, select the LONG_SLIT_K mode and long_slit_0.6
- Select DIT = 5, NDIT = 1
- wait for a couple of frames to be displayed then on the RTD, store a fixed pattern, and enable Fixed pattern
- From the Calibration Unit, switch the Xe and Ne lamps ON
- Figure shows a rotated 90 degrees Ks spectrum when the filter is well aligned, and figure shows the horizontal cut of the Xe and Ne lines.
- here can be find images of Arcs spectra taken with the J, and H filters (inc. line IDs) for comparison
- If the lines are aligned then exit procedure
- If the lines are not aligned re-iterate step 1 and 2, and CALL for HELP
B- The filter wheel is NOT ONLINE
from
the ICS Motor Device panel, put the wheel OFF, then STANDBY
1- Diagnostic
- ASSUME that you don't know what is the position of the wheel!
- On RTD open a statistics box, and move it to a relatively hot pixel free area and close to one edge.
- look at the counts: e.g. in the Ks filter the average counts should be around 2000 in morning, and 3000 in the afternoon: Ks is very sensitive to the thermal background.
2 - Search and Align
- if the counts are below 1000, or if the detector is saturating: the filter wheel is badly mis-aligned
- from the ICS control panel, open the Handset tool
- set the function to filter 1
- set the speed to 500
- use < and > to move the wheel, and the STOP button to stop the wheel from moving. Wait always for a few frames to be displayed to make sure that the function has stopped.
- Once you have found the Ks filter, reduce the speed to 50 and maximize the count level within the statistics box. Because the statistics box is close to the edge it will be sensitive to the "vigneting" due to the edge of the filter holder.
- Do a Fall Back Initialisation at the Ks position.
3- Follow step 2of the procedure A without iterating
If you cannot align in this case: CALL for HELP
Report error messages and circonstance in Remedy!