Procedure: |
SOFI grism wheel: observational strategy and recovery procedures
|
| Keywords: |
SofI - Grism Wheel - Observations - Recovery |
| Prepared/Revised: |
psinclai 2004-10-4 |
Reviewed: |
|
Released: |
epompei 2004-10-19 |
NEWS FLASH!
A software modificaion (thanks Ismo!) is taking care of the step momtion
of the grism wheel so during the operations it is not necessary to move
the wheel by hand anymore. To monitor the wheel progres:
"The little monitoring tool was also installed: isiMonitor. You can start
it from a xterm. It will start 2 plots: one for encoder position of
grims wheel and one for speed of grism wheel.
Encoder position plot should always be increasing (expect when zero is
reached). The plot is configured to have all circular range of grism
wheel, so line should go down only when upper side of the plot is
reached and then start increasing again.
The speed plot should be always greater or equal than zero.
In case you detect negative movements, you have to call software support
ASAP to recover old version. And after version is recovered, you have to
manually restrict movements again only max. 3 steps for grism wheel (and
possible aligning the grism wheel is needed)
Please, note down starting and target positions, and template name in
case of failure."
Grism wheel recovery
Currently, the grism wheel suffers from a mechanical problem that prevents a smooth movement in
the reverse direction (decreasing encoder number), which results in lost steps on the stepper motor,
and consequently, loosing the encoder value calibration. In slow speed (below 1000), the wheel does
not move at all in the reverse direction. Two setup parameter changes have made possible to use
the wheel if the user is careful. One change was to the direction of the anti-backlash compensation which
controls from which side the wheel approaches to the final position. By reversing that value, the wheel now
approaches always into the positive direction, i.e going from position 1 (wollaston 88611) to position 8 (open, 57627).
To check the wheel alignment,you can send it to the Wollaston position (88611) which
will move the wheel in + direction (from open) and will allow to check the alignment if the JG Mask is also set.
Anyway, when the wheel is not in use, it should be left in open position and Simulation mode to prevent
accidental movements.
It is also important to remember that the wheel should not move by more than three steps each
movement, i.e if you want go from wollaston to open, go first from wollaston to FE, then from FE to -ER,
then from -ER to open. Similarly from GR to GB, you have to go to -ER first, then to wollaston and
finally, at the third set-up, to GB.
IMPORTANT THINGS TO REMEMBER:
- When the grism wheel is not needed, place it in open,
then put it in simulation following Procedure B
- When using the spectroscopic mode ALWAYS follow Procedure A
to take grism wheel out of simulation, then follow the
Observational Strategy
- NEVER, EVER try to move the wheel by more than three steps
at a time. Doing so will make the motor move in the reverse
direction, the init will be lost and you'll find yourself in a downtime mess.
- If the Grism wheel position is lost and you don't know where it is, do the following:
- Put the JG mask and remove both filters, use LF. Ensure you have some background light comming in.
- Open the ICS handset and start moving the grism wheel in the positive (right) direction at a speed of 2000
and look at the RTD until you see the Wollaston (the JG spots will be duplicated). Stop immediately and try
to aproach the aligned position slowly. Use low speeds values (down to 1 enc/sec if required) until you have
a perfect vertical aligment of all spots (use auto cuts in RTD to ensure a good image).
- if you overshoot, you can move the wheel backwards for 1 or 2 seconds in speed 2000, or until you are clrealy
on the left side of the required position. Try the alignment again.
- If, by some reason, you cannot position the wheel correctly, and you need it
calibrated CALL IMMEDIATELY LIN (beeper 36),
without attempting anything else.
- if you need to align any slit, check that grism wheel is in open position and in simulation.
- If you reboot the SOFI LCU, FOLLOW Procedure C
at the end of the page.
By not doing so, the wheels will not be properly initialized
Procedure to follow if the grism wheel was in simulation (Procedure A)
- a. Open the ICS Motor Control by clicking the wheel name in ICS Control
- b. Click on the Off button in front of Simulation.
- c. The function will go to Stand-By: Press ONLINE (Init will be issued)
- d. The encoder will be loaded with the position value used last time a FB init
value was entered (it should be:"open" or 57627).
- e. Close the panel and setup the ICS with:
- Slit JG mask
- Filter 1: open
- Filter 2:open
- Grism: Wollaston (one step only is required 8-->1)
- Objective: LF
- f. You should see now in the RTD the image of the JG Mask with all spots
duplicated and perfectly aligned vertically.
If the spots are not perfectly aligned, follow the manual procedure -
RECOVERY!:
- Assume as reference position the wollaston
- Open the ICS tool handset
- Ensure the wheel is some encoder units before the correct position,
i.e., move it backwardws for a few seconds using speed 2000
- Move now the wheel toward positive direction (right arrow) using
as slow speed as required (slow approach, use any value needed, down to 1 enc. step/sec) until
the wheel is perfectly aligned.
- If you overhsoot the position, you should go backward again and try once more
- Once the wheel is aligned, do the FB init now with the position of the
Wollaston (88611) and press H/W init. Now the wheel should be aligned
- g. To use the grism wheel please do manually set-it up to the required
position, by taking care of not moving it by more than 3 positions at a time
in the + direction (i.e. from top to bottom in the positions list in
the ICS panel: Wollaston=1, open=8).
Procedure to follow if the grism wheel is not needed any longer (Procedure B)
- a. Move it to OPEN position (by going in steps of 3 positions maximum
at a time in the + direction or downwards in the list)
- b. Open the ICS Motor Control by clicking the wheel name in ICS Control
- c. Do a FB init at the open position (57627), then press HW init (it loads the open position
in the instrument database).
- d. Click on the On button in front of Simulation
- e. The function will go to Stand-By: Press ONLINE The wheel will not move anymore. Close the panel
- f. Set the Grism wheel to position Open if it is not already there to allow the user to know it is actually there.
- g. When you need to use it again, go to the previous procedure above (Procedure A)
Observational strategy
This is a description of the observation strategy to be used for
SOFI spectroscopic runs
IMPORTANT TO REMEMBER!
- The wheel can move only in positive directions, so from wollaston
to open position (1 to 8) in steps not bigger than three.
So if you wish to move from GRB to open, first go to HR, then
go to open.
- Try to optimize the observations: going from grism B to grism R
is fine, since you move one step in positive direction; going from
grism R to grims B will make you do a full turn of the wheel.
- Don't run calob in the morning if you are in charge of a
spectroscopic run. Do all calibrations in the afternoon.
- If the grism wheel looses its alignment during the run, it
will not be possible to reproduce the wavelength calibration,
so all arcs should be taken together with the object
Now let's go to real observations
Fetch the OB, put a pause between the acquisition image and the first
science template
Once the object is centered on the slit, manually set-up the grism
wheel in the required position: REMEMBER, once more, three steps per time
and ONLY in positive direction
Continue the science exposure
If the OB contains more than one science template, put a pause, then
change the grism manually. After this is done, continue through the
end of the OB.
Appendix: SOFI LCU reboot (Procedure C)
What to do in case of LCU reboot:
- ICS will stay in OFF, so before going to ONLINE:
- LOAD the database from ICS control panel ---> Configuration tool, for
the "isiInstrumentSOFI.cfg" file. This is needed because the BACKUP OFFSET must
be loaded
- Open "Motor Control Panel" (for example click on the slit wheel)
- To do a toggle: Normal --> Backup --> Normal --> BAckup for both
wheels (slit wheel only)
- ICS to ONLINE
- All wheels SHOULD be initialized
- set Slit_0.6, Ks, Open, Open (and in simulation), LF and check alignment
Appendix: SOFI FB Init (Procedure D)
How to do a FB (fall-back) Init?
- Open the ICS motor controil panel (unless it is already opened)
- Set the motor to OFF (the big upper left button of the group from
six similar buttons)
- Click on the "FALLBAKC INIT" button; this will bring up another
window - Fallback Init Launcher
- Write the encoder value corresponding to the current position in
the window; the encoder values for all positions of the current
wheel can be seen from the Positions Tables (it can be opened by
clicking on the button with the same name just above the button
for the FB Init Launcher)
- press on the "F/B Init" button in the F/B Init launcher window;
wait until the wondow displays "OK"
- press on the "Restore" button in the F/B Init launcher window
- press on "Standby" button in the ICS motor controil panel; the
color on the field above the wheel changes from red to green
- press on "Online" button in the ICS motor controil panel; the
color on the field above the wheel changes from red to gray