SciOp Procedure:

SOFI grism wheel: observational strategy and recovery procedures

Keywords: SofI - Grism Wheel - Observations - Recovery
Prepared/Revised: psinclai 2004-10-4 Reviewed: Released: epompei 2004-10-19
 

NEWS FLASH!

A software modificaion (thanks Ismo!) is taking care of the step momtion of the grism wheel so during the operations it is not necessary to move the wheel by hand anymore. To monitor the wheel progres:
"The little monitoring tool was also installed: isiMonitor. You can start it from a xterm. It will start 2 plots: one for encoder position of grims wheel and one for speed of grism wheel.
Encoder position plot should always be increasing (expect when zero is reached). The plot is configured to have all circular range of grism wheel, so line should go down only when upper side of the plot is reached and then start increasing again.
The speed plot should be always greater or equal than zero.
In case you detect negative movements, you have to call software support ASAP to recover old version. And after version is recovered, you have to manually restrict movements again only max. 3 steps for grism wheel (and possible aligning the grism wheel is needed) Please, note down starting and target positions, and template name in case of failure."

Grism wheel recovery

Currently, the grism wheel suffers from a mechanical problem that prevents a smooth movement in the reverse direction (decreasing encoder number), which results in lost steps on the stepper motor, and consequently, loosing the encoder value calibration. In slow speed (below 1000), the wheel does not move at all in the reverse direction. Two setup parameter changes have made possible to use the wheel if the user is careful. One change was to the direction of the anti-backlash compensation which controls from which side the wheel approaches to the final position. By reversing that value, the wheel now approaches always into the positive direction, i.e going from position 1 (wollaston 88611) to position 8 (open, 57627). To check the wheel alignment,you can send it to the Wollaston position (88611) which will move the wheel in + direction (from open) and will allow to check the alignment if the JG Mask is also set. Anyway, when the wheel is not in use, it should be left in open position and Simulation mode to prevent accidental movements. It is also important to remember that the wheel should not move by more than three steps each movement, i.e if you want go from wollaston to open, go first from wollaston to FE, then from FE to -ER, then from -ER to open. Similarly from GR to GB, you have to go to -ER first, then to wollaston and finally, at the third set-up, to GB.



IMPORTANT THINGS TO REMEMBER:



Procedure to follow if the grism wheel was in simulation (Procedure A)



Procedure to follow if the grism wheel is not needed any longer (Procedure B)



Observational strategy

This is a description of the observation strategy to be used for SOFI spectroscopic runs

IMPORTANT TO REMEMBER!

Now let's go to real observations

  • Fetch the OB, put a pause between the acquisition image and the first science template
  • Once the object is centered on the slit, manually set-up the grism wheel in the required position: REMEMBER, once more, three steps per time and ONLY in positive direction
  • Continue the science exposure
  • If the OB contains more than one science template, put a pause, then change the grism manually. After this is done, continue through the end of the OB.
  •  

    Appendix: SOFI LCU reboot (Procedure C)

    What to do in case of LCU reboot:

    Appendix: SOFI FB Init (Procedure D)

    How to do a FB (fall-back) Init?

     