*** MIDAS logfile (05SEPpl1.2) *** Sa, 09 Feb 2008 14:11:30 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 09 Feb 2008 14:11:44 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 080209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.24 stddev = 1.95 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_080209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1600.6 1600.2 787.3 0.13 0.12 2.03 2 4.34 3189.4 3189.4 1603.3 -0.17 -0.18 1.99 3 8.67 6382.7 6383.0 3171.7 -0.08 -0.08 2.01 4 13.00 9571.7 9572.4 4724.5 -0.10 -0.09 2.03 5 19.51 14362.9 14364.6 7128.2 -0.06 -0.05 2.01 6 26.01 19157.5 19158.4 9450.8 -0.02 -0.01 2.03 7 32.51 23955.6 23956.2 11670.3 0.02 0.02 2.05 8 43.34 31962.6 31963.2 15397.2 0.09 0.09 2.08 9 54.17 39951.0 39952.0 18996.0 0.09 0.09 2.10 10 65.01 47944.0 47945.5 22531.7 0.09 0.09 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3401 to 101890 e- Average count rate measured: 737 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.14 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay