*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Jan 2008 14:33:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Jan 2008 14:34:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 080110 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.34 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_080110 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1603.5 1603.3 785.9 0.11 0.09 2.04 2 4.34 3198.6 3199.2 1596.6 -0.13 -0.12 2.00 3 8.67 6397.2 6397.4 3177.5 -0.11 -0.11 2.01 4 13.00 9598.9 9599.0 4779.4 -0.07 -0.08 2.01 5 19.51 14400.5 14401.3 7068.2 -0.06 -0.06 2.04 6 26.01 19208.9 19211.1 9501.0 -0.01 -0.01 2.02 7 32.51 24009.4 24016.8 11605.7 -0.01 0.01 2.07 8 43.34 32054.0 32058.1 15359.3 0.12 0.12 2.09 9 54.17 40031.7 40036.5 18971.1 0.03 0.03 2.11 10 65.01 48087.3 48090.3 22392.5 0.13 0.13 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3434 to 102985 e- Average count rate measured: 739 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.19 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay