*** MIDAS logfile (05SEPpl1.2) *** Th, 27 Dec 2007 15:49:53 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 27 Dec 2007 15:50:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 071227 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.88 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_071227 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1598.2 1598.0 791.3 0.11 0.13 2.02 2 4.34 3185.6 3186.1 1608.6 -0.19 -0.16 1.98 3 8.67 6376.0 6376.0 3222.3 -0.07 -0.07 1.98 4 13.00 9561.9 9561.5 4768.7 -0.09 -0.09 2.01 5 19.51 14347.2 14348.0 7088.7 -0.06 -0.05 2.02 6 26.01 19137.6 19137.6 9379.5 -0.01 -0.01 2.04 7 32.51 23929.5 23929.8 11679.8 0.03 0.02 2.05 8 43.34 31927.0 31926.2 15412.0 0.09 0.08 2.07 9 54.17 39911.1 39911.6 18933.4 0.10 0.09 2.11 10 65.01 47883.6 47885.9 22460.4 0.08 0.08 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3405 to 102031 e- Average count rate measured: 736 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.17 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay