*** MIDAS logfile (05SEPpl1.2) *** Th, 20 Dec 2007 16:50:26 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 20 Dec 2007 16:50:40 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 071220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.70 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_071220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1599.1 1599.3 796.7 0.12 0.13 2.01 2 4.34 3187.3 3187.4 1587.2 -0.18 -0.18 2.01 3 8.67 6378.0 6378.4 3181.2 -0.08 -0.08 2.00 4 13.00 9565.9 9566.7 4774.4 -0.09 -0.09 2.00 5 19.51 14352.9 14353.2 7059.2 -0.06 -0.06 2.03 6 26.01 19145.8 19145.9 9271.7 -0.01 -0.01 2.06 7 32.51 23938.0 23939.0 11703.0 0.02 0.02 2.05 8 43.34 31942.3 31944.5 15220.0 0.10 0.10 2.10 9 54.17 39923.5 39921.0 18843.0 0.09 0.08 2.12 10 65.01 47910.9 47911.2 22043.8 0.09 0.09 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3457 to 103565 e- Average count rate measured: 736 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.26 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay