*** MIDAS logfile (05SEPpl1.2) *** We, 12 Dec 2007 13:50:14 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 12 Dec 2007 13:50:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 071212 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.91 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_071212 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1590.1 1590.2 790.2 0.12 0.13 2.01 2 4.34 3169.4 3168.9 1614.3 -0.19 -0.19 1.96 3 8.67 6343.8 6344.2 3170.3 -0.07 -0.06 2.00 4 13.00 9511.9 9510.9 4769.7 -0.10 -0.11 1.99 5 19.50 14274.0 14273.1 7009.6 -0.06 -0.06 2.04 6 26.01 19040.4 19039.5 9338.7 -0.01 -0.01 2.04 7 32.51 23809.3 23808.2 11564.1 0.03 0.03 2.06 8 43.34 31762.3 31761.1 15455.0 0.09 0.08 2.06 9 54.17 39714.0 39712.4 18852.4 0.11 0.11 2.11 10 65.01 47638.7 47635.5 22199.5 0.08 0.07 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3396 to 101733 e- Average count rate measured: 732 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.18 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay