*** MIDAS logfile (05SEPpl1.2) *** Su, 09 Dec 2007 14:54:31 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 09 Dec 2007 14:54:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 071209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.07 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_071209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1591.8 1591.9 788.3 0.13 0.14 2.02 2 4.34 3172.6 3172.4 1583.3 -0.18 -0.19 2.00 3 8.67 6350.9 6350.3 3159.0 -0.07 -0.08 2.01 4 13.00 9523.6 9523.6 4693.3 -0.10 -0.10 2.03 5 19.51 14290.1 14291.2 7022.1 -0.06 -0.06 2.04 6 26.01 19063.0 19061.8 9236.3 -0.01 -0.02 2.06 7 32.51 23836.6 23837.3 11592.2 0.02 0.03 2.06 8 43.34 31803.4 31803.1 15275.8 0.09 0.09 2.08 9 54.17 39754.5 39757.4 18907.2 0.09 0.10 2.10 10 65.01 47702.9 47703.0 22384.6 0.09 0.09 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3386 to 101458 e- Average count rate measured: 733 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.17 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay