*** MIDAS logfile (05SEPpl1.2) *** Sa, 01 Dec 2007 13:38:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 01 Dec 2007 13:38:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 071201 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.05 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_071201 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1597.2 1597.2 800.4 0.11 0.11 2.00 2 4.34 3183.0 3182.3 1576.9 -0.19 -0.20 2.02 3 8.67 6369.2 6370.3 3182.9 -0.08 -0.06 2.00 4 13.00 9551.4 9551.8 4766.0 -0.10 -0.09 2.00 5 19.51 14332.5 14333.2 7113.3 -0.05 -0.05 2.01 6 26.01 19116.5 19117.3 9298.9 -0.01 -0.01 2.06 7 32.51 23905.9 23905.4 11736.6 0.03 0.03 2.04 8 43.34 31892.7 31892.6 15198.8 0.09 0.09 2.10 9 54.17 39870.8 39869.9 18579.6 0.11 0.10 2.15 10 65.01 47832.9 47831.9 22111.8 0.08 0.08 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3459 to 103594 e- Average count rate measured: 735 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.27 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay