*** MIDAS logfile (05SEPpl1.2) *** We, 14 Nov 2007 16:43:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 14 Nov 2007 16:43:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 071114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.30 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_071114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1600.7 1600.1 796.7 0.14 0.14 2.01 2 4.34 3185.7 3186.7 1574.9 -0.23 -0.21 2.02 3 8.67 6378.0 6377.9 3130.5 -0.07 -0.07 2.04 4 13.00 9562.1 9561.4 4790.3 -0.11 -0.11 2.00 5 19.51 14351.2 14348.6 7034.9 -0.04 -0.06 2.04 6 26.01 19139.3 19138.6 9240.5 -0.01 -0.01 2.07 7 32.51 23937.0 23936.2 11695.3 0.04 0.04 2.05 8 43.34 31928.7 31925.8 15221.0 0.09 0.08 2.10 9 54.17 39930.1 39925.6 18864.7 0.14 0.13 2.12 10 65.01 47880.0 47881.8 22294.4 0.06 0.07 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3431 to 102649 e- Average count rate measured: 736 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.23 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay