*** MIDAS logfile (05SEPpl1.2) *** Sa, 03 Nov 2007 15:05:01 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 03 Nov 2007 15:05:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 071103 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.19 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_071103 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1600.3 1600.7 790.4 0.12 0.12 2.02 2 4.34 3189.6 3189.3 1582.6 -0.19 -0.19 2.02 3 8.67 6384.0 6383.5 3212.1 -0.07 -0.08 1.99 4 13.00 9573.0 9572.5 4751.8 -0.10 -0.10 2.01 5 19.51 14364.8 14364.2 7045.1 -0.06 -0.06 2.04 6 26.01 19161.8 19160.9 9371.3 -0.01 -0.01 2.04 7 32.51 23960.0 23959.1 11871.6 0.03 0.02 2.02 8 43.34 31967.1 31966.0 15355.9 0.09 0.09 2.08 9 54.17 39962.9 39965.4 18963.8 0.10 0.11 2.11 10 65.01 47947.4 47948.9 22254.4 0.09 0.09 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3428 to 102692 e- Average count rate measured: 737 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.24 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay