*** MIDAS logfile (05SEPpl1.2) *** We, 26 Sep 2007 17:45:38 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 26 Sep 2007 17:45:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070926 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.69 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070926 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1603.3 1603.8 789.7 0.13 0.13 2.03 2 4.34 3195.4 3195.1 1588.6 -0.18 -0.18 2.01 3 8.67 6395.8 6395.3 3107.1 -0.07 -0.07 2.06 4 13.00 9590.3 9590.4 4785.8 -0.10 -0.09 2.00 5 19.51 14389.7 14390.5 7129.5 -0.06 -0.06 2.02 6 26.01 19194.2 19193.5 9454.7 -0.01 -0.02 2.03 7 32.51 24001.8 24002.3 11588.3 0.03 0.03 2.07 8 43.34 32021.7 32021.7 15305.7 0.09 0.09 2.09 9 54.17 40030.0 40030.8 19287.4 0.10 0.10 2.08 10 65.01 48028.1 48026.2 22465.3 0.08 0.08 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3405 to 101967 e- Average count rate measured: 738 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.19 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay