*** MIDAS logfile (05SEPpl1.2) *** We, 19 Sep 2007 18:18:30 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 19 Sep 2007 18:18:44 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070919 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.62 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070919 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1599.7 1599.8 788.7 0.11 0.12 2.03 2 4.34 3188.1 3187.7 1593.5 -0.19 -0.20 2.00 3 8.67 6380.8 6380.8 3204.3 -0.08 -0.08 1.99 4 13.00 9569.0 9569.7 4748.2 -0.09 -0.09 2.02 5 19.51 14358.3 14358.1 7138.0 -0.06 -0.06 2.01 6 26.01 19152.9 19153.5 9326.0 -0.01 -0.01 2.05 7 32.51 23948.2 23950.1 12015.9 0.02 0.03 1.99 8 43.34 31954.3 31954.6 15417.1 0.10 0.10 2.07 9 54.17 39941.3 39941.4 18798.9 0.10 0.09 2.12 10 65.01 47932.4 47929.8 22387.5 0.10 0.10 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3421 to 102489 e- Average count rate measured: 737 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.03 e-/ADU RON = 4.21 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay