*** MIDAS logfile (05SEPpl1.2) *** We, 29 Aug 2007 17:57:00 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 29 Aug 2007 17:57:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070829 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.76 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070829 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1606.3 1606.2 804.9 0.12 0.12 2.00 2 4.34 3200.9 3201.7 1598.4 -0.20 -0.18 2.00 3 8.67 6406.4 6407.3 3199.4 -0.09 -0.07 2.00 4 13.00 9608.0 9608.3 4724.7 -0.10 -0.09 2.03 5 19.51 14417.3 14416.5 7084.2 -0.06 -0.06 2.04 6 26.01 19231.0 19232.0 9433.5 -0.01 -0.01 2.04 7 32.51 24046.4 24047.3 11643.0 0.02 0.02 2.07 8 43.34 32086.1 32080.9 15407.9 0.10 0.08 2.08 9 54.17 40106.4 40106.7 19034.7 0.10 0.10 2.11 10 65.01 48130.9 48126.8 22518.3 0.11 0.10 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3426 to 102670 e- Average count rate measured: 740 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay