*** MIDAS logfile (05SEPpl1.2) *** Mo, 27 Aug 2007 15:10:05 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 27 Aug 2007 15:10:19 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070827 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.82 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070827 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1600.3 1600.2 785.6 0.15 0.15 2.04 2 4.34 3187.0 3187.9 1586.1 -0.22 -0.20 2.01 3 8.67 6381.4 6381.9 3197.4 -0.07 -0.06 2.00 4 13.00 9568.2 9568.2 4824.7 -0.11 -0.11 1.98 5 19.51 14358.6 14358.5 7079.2 -0.06 -0.06 2.03 6 26.01 19153.1 19152.8 9360.8 -0.01 -0.01 2.05 7 32.51 23955.0 23953.2 11749.4 0.05 0.04 2.04 8 43.34 31950.3 31948.4 15343.1 0.08 0.08 2.08 9 54.17 39950.2 39948.5 18866.0 0.12 0.11 2.12 10 65.01 47919.4 47917.5 22522.5 0.07 0.07 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3412 to 102171 e- Average count rate measured: 737 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.22 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay