*** MIDAS logfile (05SEPpl1.2) *** Th, 09 Aug 2007 15:53:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 09 Aug 2007 15:53:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070809 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.74 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070809 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1617.0 1617.0 815.1 0.12 0.11 1.98 2 4.34 3224.4 3225.3 1605.3 -0.15 -0.14 2.01 3 8.67 6451.2 6451.5 3205.4 -0.10 -0.10 2.01 4 13.00 9677.7 9679.2 4812.2 -0.08 -0.08 2.01 5 19.51 14519.5 14520.4 7202.3 -0.06 -0.06 2.02 6 26.01 19368.5 19371.2 9397.3 -0.01 -0.01 2.06 7 32.51 24216.2 24216.7 11852.3 0.01 0.00 2.04 8 43.34 32318.8 32322.0 15492.1 0.11 0.11 2.09 9 54.17 40378.6 40382.5 19035.7 0.06 0.06 2.12 10 65.01 48482.0 48485.8 22668.3 0.12 0.12 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3461 to 103764 e- Average count rate measured: 745 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.24 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay