*** MIDAS logfile (05SEPpl1.2) *** Th, 02 Aug 2007 16:15:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 02 Aug 2007 16:15:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070802 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.88 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070802 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1608.3 1608.6 806.0 0.12 0.12 2.00 2 4.34 3204.4 3204.0 1597.2 -0.21 -0.22 2.01 3 8.67 6414.7 6415.3 3196.8 -0.07 -0.07 2.01 4 13.00 9620.9 9619.3 4718.1 -0.09 -0.10 2.04 5 19.51 14434.1 14435.9 7087.3 -0.06 -0.05 2.04 6 26.01 19253.9 19254.2 9442.3 -0.01 -0.01 2.04 7 32.51 24077.0 24078.6 11441.9 0.03 0.04 2.10 8 43.34 32121.1 32121.3 15373.2 0.09 0.09 2.09 9 54.17 40157.1 40160.8 18960.4 0.11 0.11 2.12 10 65.01 48178.9 48179.2 22485.4 0.09 0.09 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3447 to 103240 e- Average count rate measured: 740 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 4.25 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay