*** MIDAS logfile (05SEPpl1.2) *** We, 25 Jul 2007 15:22:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 25 Jul 2007 15:22:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070725 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.06 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070725 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1607.6 1607.6 811.4 11.26 25.12 1.98 2 4.34 3200.4 0.0 974.7 10.82 -100.00 3.28 3 8.67 6412.2 6412.6 3172.9 11.08 24.93 2.02 4 13.00 9610.8 9609.9 4701.9 11.00 24.83 2.04 5 19.51 14426.1 14424.2 7102.6 11.08 24.92 2.03 6 26.01 19238.5 19238.6 9361.7 11.11 24.97 2.06 7 32.51 24063.0 24061.7 11490.1 11.18 25.04 2.09 8 43.34 32083.4 32080.3 15276.3 11.18 25.04 2.10 9 54.17 40139.4 40137.8 18819.6 11.28 25.16 2.13 10 65.01 0.1 0.2 3.9 -100.00 -100.00 0.03 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 0 to 85607 e- Average count rate measured: 629 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.03 e-/ADU RON = 4.21 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = +18.22/-100.00 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay