*** MIDAS logfile (05SEPpl1.2) *** Fr, 15 Jun 2007 15:42:39 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 15 Jun 2007 15:42:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070615 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.53 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070615 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1631.6 1631.7 814.8 0.12 0.10 2.00 2 4.34 3252.7 3253.1 1652.8 -0.16 -0.17 1.97 3 8.67 6509.1 6509.7 3268.6 -0.08 -0.07 1.99 4 13.00 9760.7 9762.8 4821.4 -0.11 -0.09 2.02 5 19.51 14647.7 14647.1 7157.8 -0.06 -0.06 2.05 6 26.01 19536.9 19536.1 9513.4 -0.02 -0.02 2.05 7 32.51 24429.8 24429.4 11933.1 0.02 0.01 2.05 8 43.34 32603.8 32603.1 15680.9 0.11 0.11 2.08 9 54.17 40738.6 40744.0 19461.0 0.07 0.09 2.09 10 65.01 48896.4 48897.9 22979.5 0.09 0.10 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3463 to 103789 e- Average count rate measured: 751 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.18 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay