*** MIDAS logfile (05SEPpl1.2) *** Th, 07 Jun 2007 15:13:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 07 Jun 2007 15:13:57 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070607 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.80 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070607 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 7.3 7.3 7.7 -0.22 -0.31 0.94 2 4.34 14.4 14.4 12.0 -0.50 -0.23 1.20 3 8.67 29.3 29.2 20.0 1.57 1.44 1.46 4 13.00 43.6 43.6 28.7 0.99 0.93 1.52 5 19.51 64.8 64.8 36.8 0.13 0.05 1.76 6 26.01 86.0 86.0 46.9 -0.29 -0.31 1.83 7 32.51 107.3 107.3 58.1 -0.48 -0.47 1.85 8 43.34 143.1 143.1 84.3 -0.43 -0.41 1.70 9 54.17 179.1 179.1 96.0 -0.25 -0.24 1.87 10 65.01 214.4 214.5 112.3 -0.52 -0.46 1.91 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 14 to 419 e- Average count rate measured: 3 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.95 +/- 0.05 e-/ADU RON = 3.85 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.032 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.51/-0.50 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay