*** MIDAS logfile (05SEPpl1.2) *** Th, 24 May 2007 16:03:32 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 24 May 2007 16:03:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070524 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.48 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070524 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1656.5 1656.6 831.9 0.14 0.12 1.99 2 4.34 3299.4 3299.2 1629.8 -0.21 -0.22 2.02 3 8.67 6605.7 6605.6 3276.6 -0.06 -0.06 2.02 4 13.00 9903.5 9903.9 4850.3 -0.10 -0.10 2.04 5 19.51 14862.5 14861.8 7277.1 -0.05 -0.05 2.04 6 26.01 19823.9 19824.2 9664.0 -0.01 0.00 2.05 7 32.51 24791.0 24790.5 12149.2 0.04 0.04 2.04 8 43.34 33068.1 33066.7 15636.4 0.08 0.08 2.11 9 54.17 41352.8 41351.3 19431.7 0.12 0.13 2.13 10 65.01 49586.4 49584.7 22894.8 0.05 0.06 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3579 to 107122 e- Average count rate measured: 762 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.27 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay