*** MIDAS logfile (05SEPpl1.2) *** We, 02 May 2007 15:06:11 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 02 May 2007 15:06:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070502 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.36 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070502 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1668.4 1667.9 834.8 0.14 0.13 2.00 2 4.34 3323.1 3322.6 1673.1 -0.21 -0.21 1.99 3 8.67 6653.4 6652.7 3292.9 -0.06 -0.07 2.02 4 13.00 9976.7 9975.6 4945.5 -0.10 -0.10 2.02 5 19.51 14971.6 14971.3 7271.2 -0.05 -0.04 2.06 6 26.01 19969.1 19968.8 9608.9 -0.01 -0.00 2.08 7 32.51 24972.8 24971.3 12100.0 0.04 0.04 2.06 8 43.34 33310.2 33311.1 15939.5 0.08 0.09 2.09 9 54.17 41646.3 41647.7 19616.8 0.10 0.11 2.12 10 65.01 49952.2 49949.0 22886.9 0.06 0.06 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3613 to 108195 e- Average count rate measured: 768 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.26 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay