*** MIDAS logfile (05SEPpl1.2) *** Sa, 28 Apr 2007 15:52:22 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 28 Apr 2007 15:52:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070428 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.29 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070428 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1668.3 1667.2 836.3 0.12 0.12 1.99 2 4.34 3323.8 3323.3 1658.4 -0.19 -0.19 2.00 3 8.67 6653.6 6653.5 3255.8 -0.07 -0.06 2.04 4 13.00 9976.8 9976.6 4998.7 -0.10 -0.10 2.00 5 19.51 14971.7 14971.6 7380.7 -0.05 -0.05 2.03 6 26.01 19968.8 19969.6 9764.6 -0.01 -0.01 2.05 7 32.51 24973.1 24973.8 11987.4 0.04 0.04 2.08 8 43.34 33313.9 33314.3 16098.9 0.09 0.09 2.07 9 54.17 41653.3 41654.1 19713.1 0.11 0.11 2.11 10 65.01 49959.0 49957.8 23007.2 0.07 0.06 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3593 to 107616 e- Average count rate measured: 768 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.24 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay