*** MIDAS logfile (05SEPpl1.2) *** Mo, 23 Apr 2007 16:01:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 23 Apr 2007 16:01:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070423 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.22 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070423 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1674.1 1675.0 842.5 0.14 0.13 1.99 2 4.34 3336.9 3336.8 1652.6 -0.18 -0.19 2.02 3 8.67 6679.4 6678.6 3282.3 -0.07 -0.09 2.03 4 13.00 10016.5 10016.9 5033.0 -0.10 -0.09 1.99 5 19.51 15031.1 15031.1 7430.7 -0.05 -0.05 2.02 6 26.01 20048.0 20049.0 9702.3 -0.01 -0.01 2.07 7 32.51 25069.3 25070.7 12127.1 0.02 0.03 2.07 8 43.34 33448.2 33448.1 15902.3 0.09 0.09 2.10 9 54.17 41812.7 41811.8 19586.4 0.10 0.10 2.13 10 65.01 50161.1 50160.6 23086.8 0.07 0.07 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3632 to 108784 e- Average count rate measured: 771 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.26 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay