*** MIDAS logfile (05SEPpl1.2) *** We, 11 Apr 2007 14:04:09 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 11 Apr 2007 14:04:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070411 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.43 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070411 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1687.1 1687.3 844.6 0.12 0.12 2.00 2 4.34 3361.2 3361.3 1686.4 -0.21 -0.21 1.99 3 8.67 6730.5 6729.7 3328.0 -0.06 -0.07 2.02 4 13.00 10090.6 10091.5 4929.9 -0.10 -0.10 2.05 5 19.51 15144.4 15144.2 7515.5 -0.05 -0.05 2.02 6 26.01 20199.4 20199.9 9980.6 -0.01 -0.00 2.02 7 32.51 25261.5 25260.0 12132.7 0.04 0.04 2.08 8 43.34 33697.7 33696.2 16040.1 0.09 0.09 2.10 9 54.17 42129.8 42131.5 19707.1 0.11 0.12 2.14 10 65.01 50529.9 50529.7 23194.1 0.06 0.06 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3665 to 109768 e- Average count rate measured: 777 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.29 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay