*** MIDAS logfile (05SEPpl1.2) *** Sa, 24 Mar 2007 17:23:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 24 Mar 2007 17:23:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070324 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.41 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070324 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1692.8 1692.0 854.7 0.14 0.12 1.98 2 4.34 3372.0 3372.7 1696.6 -0.20 -0.18 1.99 3 8.67 6750.0 6749.6 3348.1 -0.07 -0.07 2.02 4 13.00 10121.8 10121.0 4981.6 -0.09 -0.09 2.03 5 19.51 15188.3 15187.9 7487.1 -0.05 -0.05 2.03 6 26.01 20259.2 20257.9 9869.0 -0.00 -0.01 2.05 7 32.51 25332.6 25331.5 12157.2 0.03 0.03 2.08 8 43.34 33792.5 33794.1 16452.3 0.08 0.09 2.05 9 54.17 42248.1 42247.0 20041.3 0.10 0.10 2.11 10 65.01 50678.7 50676.8 23592.7 0.06 0.06 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3615 to 108242 e- Average count rate measured: 779 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.22 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay