*** MIDAS logfile (05SEPpl1.2) *** Mo, 19 Mar 2007 14:12:35 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 19 Mar 2007 14:12:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070319 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.14 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070319 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1694.0 1694.4 856.6 0.10 0.12 1.98 2 4.34 3377.6 3377.5 1697.6 -0.16 -0.17 1.99 3 8.67 6758.8 6759.6 3349.8 -0.07 -0.06 2.02 4 13.00 10136.3 10135.7 5060.3 -0.08 -0.09 2.00 5 19.51 15208.9 15209.1 7429.3 -0.05 -0.05 2.05 6 26.01 20285.4 20287.4 9922.0 -0.01 -0.00 2.04 7 32.51 25366.6 25367.2 12252.3 0.03 0.03 2.07 8 43.34 33843.0 33842.1 16137.7 0.09 0.09 2.10 9 54.17 42301.1 42301.3 19800.2 0.09 0.09 2.14 10 65.01 50749.5 50749.9 23665.9 0.07 0.06 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3646 to 109204 e- Average count rate measured: 780 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.22 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay