*** MIDAS logfile (05SEPpl1.2) *** Su, 25 Feb 2007 15:23:36 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 25 Feb 2007 15:23:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070225 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.27 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070225 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1700.5 1700.9 866.7 0.09 0.10 1.96 2 4.34 3390.6 3389.7 1686.9 -0.17 -0.19 2.01 3 8.67 6784.6 6785.0 4242.3 -0.07 -0.07 1.60 4 13.00 10175.3 10174.9 5107.0 -0.09 -0.09 1.99 5 19.51 15267.6 15267.4 7480.0 -0.05 -0.05 2.04 6 26.01 20365.9 20366.6 9974.0 -0.00 -0.00 2.04 7 32.51 25465.7 25467.1 12377.4 0.03 0.04 2.06 8 43.34 33977.2 33976.6 16182.5 0.10 0.10 2.10 9 54.17 42465.1 42467.4 20030.1 0.09 0.09 2.12 10 65.01 50952.1 50952.2 23522.3 0.08 0.08 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3711 to 111175 e- Average count rate measured: 783 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.03 e-/ADU RON = 4.31 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay