*** MIDAS logfile (05SEPpl1.2) *** Fr, 16 Feb 2007 16:29:42 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 16 Feb 2007 16:29:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070216 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.16 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070216 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1708.1 1707.2 866.8 0.09 0.12 1.97 2 4.34 3404.5 3403.4 1700.7 -0.17 -0.16 2.00 3 8.67 6811.3 6810.9 3332.8 -0.08 -0.08 2.04 4 13.00 10216.3 10215.9 5119.4 -0.08 -0.08 2.00 5 19.51 15327.8 15328.8 7574.4 -0.05 -0.05 2.02 6 26.01 20446.3 20446.6 9929.5 0.00 -0.01 2.06 7 32.51 25565.2 25565.4 12408.4 0.03 0.02 2.06 8 43.34 34109.2 34111.7 16420.5 0.10 0.10 2.08 9 54.17 42627.9 42630.1 20131.0 0.08 0.07 2.12 10 65.01 51149.2 51151.5 23651.1 0.07 0.07 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3674 to 110057 e- Average count rate measured: 786 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.23 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay