*** MIDAS logfile (05SEPpl1.2) *** Su, 04 Feb 2007 17:25:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 04 Feb 2007 17:25:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070204 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.07 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070204 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1717.1 1716.9 865.5 0.13 0.13 1.98 2 4.34 3421.8 3421.7 1710.4 -0.17 -0.18 2.00 3 8.67 6847.3 6847.5 3372.0 -0.08 -0.08 2.03 4 13.00 10269.9 10269.2 5058.8 -0.09 -0.10 2.03 5 19.51 15409.3 15409.6 7575.1 -0.06 -0.06 2.03 6 26.01 20555.8 20555.7 10080.7 -0.00 -0.00 2.04 7 32.51 25703.3 25704.0 12483.6 0.03 0.03 2.06 8 43.34 34293.9 34293.0 16289.5 0.10 0.10 2.11 9 54.17 42857.0 42861.2 20179.0 0.08 0.09 2.12 10 65.01 51422.4 51424.7 23869.4 0.07 0.07 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3696 to 110685 e- Average count rate measured: 790 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.24 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay