*** MIDAS logfile (05SEPpl1.2) *** Su, 07 Jan 2007 15:47:52 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 07 Jan 2007 15:48:06 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070107 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.97 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070107 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1715.0 1714.9 870.5 0.10 0.13 1.97 2 4.34 3417.3 3417.1 1685.5 -0.19 -0.17 2.03 3 8.67 6837.7 6837.7 3381.6 -0.08 -0.08 2.02 4 13.00 10255.5 10254.9 5055.6 -0.09 -0.10 2.03 5 19.51 15388.3 15389.3 7461.3 -0.05 -0.05 2.06 6 26.01 20526.1 20526.9 9953.8 0.00 -0.00 2.06 7 32.51 25666.3 25668.0 12358.1 0.03 0.03 2.08 8 43.34 34244.5 34242.4 16224.3 0.10 0.09 2.11 9 54.17 42798.1 42799.2 20045.2 0.09 0.08 2.14 10 65.01 51348.4 51351.1 23493.0 0.07 0.07 2.19 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3731 to 111707 e- Average count rate measured: 789 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.02 e-/ADU RON = 4.26 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay