*** MIDAS logfile (05SEPpl1.2) *** Tu, 02 Jan 2007 12:33:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Tu, 02 Jan 2007 12:33:58 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070102 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.90 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_070102 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1709.7 1709.6 865.1 0.09 0.12 1.98 2 4.34 3409.2 3408.3 1684.1 -0.16 -0.17 2.02 3 8.67 6821.7 6822.2 3423.8 -0.08 -0.07 1.99 4 13.00 10231.1 10231.2 5073.8 -0.08 -0.08 2.02 5 19.51 15352.3 15351.0 7532.6 -0.04 -0.06 2.04 6 26.01 20476.3 20477.1 10116.6 -0.01 -0.01 2.02 7 32.51 25603.8 25605.8 12209.1 0.03 0.03 2.10 8 43.34 34161.5 34164.6 16305.1 0.10 0.10 2.10 9 54.17 42690.2 42692.8 20352.5 0.07 0.07 2.10 10 65.01 51232.1 51231.1 23599.2 0.08 0.07 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3683 to 110361 e- Average count rate measured: 788 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.24 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay