*** MIDAS logfile (05SEPpl1.2) *** Th, 21 Dec 2006 16:31:22 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 21 Dec 2006 16:31:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061221 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.41 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061221 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1704.8 1704.8 865.6 0.10 0.10 1.97 2 4.34 3398.3 3398.9 1712.0 -0.18 -0.17 1.98 3 8.67 6800.4 6800.6 3343.3 -0.08 -0.08 2.03 4 13.00 10200.9 10200.2 5023.8 -0.08 -0.09 2.03 5 19.51 15304.9 15305.3 7551.6 -0.05 -0.05 2.03 6 26.01 20415.1 20416.9 10105.6 -0.00 0.00 2.02 7 32.51 25523.7 25526.4 12199.3 0.02 0.03 2.09 8 43.34 34060.9 34059.9 16141.7 0.11 0.10 2.11 9 54.17 42563.4 42563.0 20150.0 0.08 0.07 2.11 10 65.01 51079.2 51081.1 23788.5 0.09 0.09 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3661 to 109684 e- Average count rate measured: 785 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.22 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay