*** MIDAS logfile (05SEPpl1.2) *** We, 06 Dec 2006 15:44:22 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Dec 2006 15:44:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061206 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.38 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061206 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1694.8 1695.0 846.6 0.13 0.13 2.00 2 4.34 3379.0 3379.1 1699.6 -0.17 -0.16 1.99 3 8.67 6762.0 6762.3 3323.0 -0.08 -0.08 2.03 4 13.00 10142.2 10142.1 4991.1 -0.09 -0.09 2.03 5 19.51 15218.7 15218.6 7563.1 -0.06 -0.06 2.01 6 26.01 20298.9 20300.6 9920.1 -0.01 -0.01 2.05 7 32.51 25383.7 25383.8 12269.7 0.02 0.02 2.07 8 43.34 33869.1 33870.9 16213.7 0.10 0.10 2.09 9 54.17 42327.7 42326.6 19893.5 0.08 0.07 2.13 10 65.01 50794.2 50794.9 23753.5 0.08 0.08 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3631 to 108826 e- Average count rate measured: 781 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.23 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay