*** MIDAS logfile (05SEPpl1.2) *** We, 29 Nov 2006 15:19:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 29 Nov 2006 15:20:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061129 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.52 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061129 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1704.4 1704.4 885.8 0.11 0.11 1.92 2 4.34 3396.0 3396.4 1695.3 -0.20 -0.20 2.00 3 8.67 6797.7 6797.3 3378.5 -0.07 -0.07 2.01 4 13.00 10193.5 10193.6 4992.1 -0.09 -0.09 2.04 5 19.51 15296.4 15296.1 7638.7 -0.04 -0.04 2.00 6 26.01 20401.1 20400.5 9849.5 -0.01 -0.01 2.07 7 32.51 25511.5 25512.8 12453.0 0.03 0.04 2.05 8 43.34 34036.2 34034.6 16326.6 0.10 0.09 2.08 9 54.17 42550.7 42545.1 20147.6 0.11 0.10 2.11 10 65.01 51034.6 51037.3 23945.7 0.06 0.07 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3634 to 108828 e- Average count rate measured: 785 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.20 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay