*** MIDAS logfile (05SEPpl1.2) *** Su, 26 Nov 2006 13:23:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 26 Nov 2006 13:24:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061126 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.53 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061126 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1705.3 1705.4 866.0 0.12 0.12 1.97 2 4.34 3398.5 3398.9 1704.1 -0.18 -0.17 1.99 3 8.67 6800.4 6800.7 3398.9 -0.08 -0.08 2.00 4 13.00 10198.9 10199.5 5001.5 -0.10 -0.09 2.04 5 19.51 15304.1 15303.7 7494.6 -0.06 -0.06 2.04 6 26.01 20414.8 20415.3 9926.3 -0.00 -0.00 2.06 7 32.51 25526.5 25525.3 12358.0 0.03 0.02 2.07 8 43.34 34060.1 34060.5 16409.4 0.10 0.10 2.08 9 54.17 42569.9 42569.7 20411.8 0.09 0.09 2.09 10 65.01 51077.0 51079.0 23526.9 0.08 0.08 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3656 to 109503 e- Average count rate measured: 785 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.23 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay