*** MIDAS logfile (05SEPpl1.2) *** We, 15 Nov 2006 14:35:23 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 15 Nov 2006 14:35:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061115 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.50 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061115 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1709.2 1708.0 871.4 0.15 0.11 1.96 2 4.34 3402.9 3403.7 1716.6 -0.21 -0.19 1.98 3 8.67 6812.3 6812.5 3378.2 -0.07 -0.06 2.02 4 13.00 10214.8 10214.9 5069.2 -0.10 -0.09 2.02 5 19.51 15328.5 15327.1 7602.5 -0.05 -0.05 2.02 6 26.01 20445.5 20444.9 9981.6 -0.00 0.00 2.05 7 32.51 25566.6 25565.3 12513.3 0.04 0.04 2.04 8 43.34 34104.2 34102.7 16384.8 0.09 0.09 2.08 9 54.17 42640.6 42636.6 19935.3 0.11 0.11 2.14 10 65.01 51135.7 51135.7 23695.6 0.05 0.05 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3687 to 110335 e- Average count rate measured: 786 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.25 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay