*** MIDAS logfile (05SEPpl1.2) *** Fr, 10 Nov 2006 16:20:54 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 10 Nov 2006 16:21:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061110 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.41 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061110 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1709.6 1710.4 860.9 0.12 0.14 1.99 2 4.34 3406.6 3406.6 1733.7 -0.19 -0.19 1.96 3 8.67 6817.0 6817.3 3387.9 -0.07 -0.08 2.01 4 13.00 10223.2 10224.2 5136.1 -0.09 -0.09 1.99 5 19.51 15340.3 15340.7 7609.9 -0.05 -0.05 2.02 6 26.01 20460.9 20462.7 10062.4 -0.01 -0.00 2.03 7 32.51 25585.6 25587.2 12315.8 0.03 0.03 2.08 8 43.34 34133.4 34134.2 16236.6 0.09 0.09 2.10 9 54.17 42670.7 42670.1 20106.6 0.10 0.10 2.12 10 65.01 51183.6 51184.1 23772.8 0.06 0.06 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3686 to 110340 e- Average count rate measured: 787 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.26 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay