*** MIDAS logfile (05SEPpl1.2) *** Sa, 04 Nov 2006 12:56:47 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 04 Nov 2006 12:57:01 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061104 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.77 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1698.1 1699.0 877.5 0.10 0.13 1.94 2 4.34 3385.4 3385.2 1697.0 -0.18 -0.19 1.99 3 8.67 6775.5 6775.4 3367.8 -0.07 -0.07 2.01 4 13.00 10158.7 10159.3 5042.4 -0.10 -0.10 2.01 5 19.51 15247.1 15246.4 7384.6 -0.04 -0.05 2.06 6 26.01 20336.2 20337.0 9852.3 -0.00 -0.00 2.06 7 32.51 25430.3 25431.2 12182.8 0.04 0.04 2.09 8 43.34 33925.5 33923.4 16306.9 0.09 0.08 2.08 9 54.17 42413.5 42412.0 19951.6 0.11 0.10 2.13 10 65.01 50872.3 50872.7 23495.1 0.06 0.06 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3665 to 109755 e- Average count rate measured: 782 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.25 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay