*** MIDAS logfile (05SEPpl1.2) *** We, 25 Oct 2006 15:48:37 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 25 Oct 2006 15:48:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061025 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.86 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061025 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1699.3 1699.6 859.0 0.11 0.13 1.98 2 4.34 3386.6 3386.8 1692.7 -0.18 -0.18 2.00 3 8.67 6777.7 6778.1 3348.5 -0.07 -0.07 2.02 4 13.00 10163.1 10164.2 5066.7 -0.10 -0.09 2.01 5 19.51 15252.0 15251.7 7484.3 -0.05 -0.05 2.04 6 26.01 20344.3 20343.0 9862.3 -0.00 -0.01 2.06 7 32.51 25439.0 25438.6 12267.2 0.03 0.03 2.07 8 43.34 33939.1 33938.6 15909.8 0.09 0.09 2.13 9 54.17 42421.8 42420.3 19920.8 0.09 0.09 2.13 10 65.01 50897.7 50899.9 23671.6 0.08 0.08 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3669 to 109881 e- Average count rate measured: 782 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.01 e-/ADU RON = 4.26 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay