*** MIDAS logfile (05SEPpl1.2) *** We, 18 Oct 2006 13:07:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 18 Oct 2006 13:07:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061018 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.06 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1710.1 1710.0 871.8 0.11 0.10 1.96 2 4.34 3407.2 3407.8 1682.5 -0.20 -0.19 2.03 3 8.67 6820.1 6820.2 3447.9 -0.07 -0.07 1.98 4 13.00 10226.6 10227.2 5045.8 -0.09 -0.09 2.03 5 19.51 15345.1 15345.2 7604.8 -0.05 -0.05 2.02 6 26.01 20469.3 20468.2 10053.2 0.00 -0.01 2.04 7 32.51 25595.8 25596.0 12481.6 0.04 0.04 2.05 8 43.34 34146.8 34145.6 16053.0 0.10 0.09 2.13 9 54.17 42682.2 42684.4 20204.9 0.10 0.10 2.11 10 65.01 51201.2 51203.6 23455.8 0.06 0.07 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3713 to 111172 e- Average count rate measured: 787 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.03 e-/ADU RON = 4.29 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay