*** MIDAS logfile (05SEPpl1.2) *** Th, 12 Oct 2006 17:36:14 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 12 Oct 2006 17:36:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061012 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.88 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061012 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1706.2 1706.9 870.9 0.11 0.13 1.96 2 4.34 3401.4 3401.2 1705.6 -0.16 -0.18 1.99 3 8.67 6805.6 6806.2 3345.6 -0.08 -0.08 2.03 4 13.00 10207.7 10207.6 5054.0 -0.09 -0.09 2.02 5 19.51 15315.6 15317.3 7525.7 -0.06 -0.05 2.04 6 26.01 20431.0 20430.5 9972.0 -0.00 -0.01 2.05 7 32.51 25545.8 25546.4 12368.1 0.03 0.03 2.07 8 43.34 34084.2 34085.3 16141.7 0.10 0.10 2.11 9 54.17 42595.5 42596.4 20023.8 0.08 0.08 2.13 10 65.01 51114.3 51116.7 23320.3 0.08 0.08 2.19 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3717 to 111322 e- Average count rate measured: 786 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.02 e-/ADU RON = 4.28 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay