*** MIDAS logfile (05SEPpl1.2) *** We, 04 Oct 2006 15:22:28 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 04 Oct 2006 15:22:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061004 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.95 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_061004 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1706.3 1706.1 869.2 0.13 0.12 1.96 2 4.34 3399.6 3399.4 1686.3 -0.18 -0.18 2.02 3 8.67 6803.9 6804.1 3371.0 -0.08 -0.07 2.02 4 13.00 10204.3 10203.6 5042.2 -0.09 -0.09 2.02 5 19.51 15311.4 15311.6 7572.2 -0.05 -0.05 2.02 6 26.01 20422.6 20422.8 10119.0 -0.01 -0.01 2.02 7 32.51 25537.4 25537.7 12463.4 0.03 0.03 2.05 8 43.34 34072.3 34072.1 16377.8 0.10 0.10 2.08 9 54.17 42583.7 42585.5 20368.9 0.08 0.09 2.09 10 65.01 51090.2 51090.4 23730.2 0.06 0.07 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3644 to 109122 e- Average count rate measured: 785 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.21 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay