*** MIDAS logfile (05SEPpl1.2) *** We, 27 Sep 2006 19:57:28 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 27 Sep 2006 19:57:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060927 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.93 stddev = 2.17 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060927 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1711.0 1711.0 871.1 0.10 0.09 1.96 2 4.34 3410.9 3410.9 1713.6 -0.13 -0.14 1.99 3 8.67 6821.0 6820.7 3407.9 -0.09 -0.10 2.00 4 13.00 10228.6 10236.1 5137.3 -0.11 -0.04 1.99 5 19.51 15347.1 15347.0 7566.6 -0.08 -0.08 2.03 6 26.01 20470.9 20471.1 9903.4 -0.03 -0.03 2.07 7 32.51 25598.2 25597.8 12472.6 0.01 0.00 2.05 8 43.34 34175.1 34176.7 16306.6 0.14 0.14 2.10 9 54.17 42685.1 42686.4 19942.8 0.07 0.06 2.14 10 65.01 51245.2 51245.2 23633.6 0.11 0.11 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3710 to 111119 e- Average count rate measured: 787 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.71 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay