*** MIDAS logfile (05SEPpl1.2) *** We, 13 Sep 2006 15:18:22 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 13 Sep 2006 15:18:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060913 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.08 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060913 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1711.5 1712.3 860.8 0.14 0.14 1.99 2 4.34 3409.8 3409.7 1721.1 -0.20 -0.20 1.98 3 8.67 6824.8 6826.0 3387.2 -0.07 -0.06 2.01 4 13.00 10233.9 10234.1 5100.2 -0.10 -0.10 2.01 5 19.51 15357.6 15358.1 7624.0 -0.05 -0.05 2.01 6 26.01 20484.4 20484.8 9994.8 -0.01 -0.00 2.05 7 32.51 25616.9 25617.6 12381.5 0.04 0.04 2.07 8 43.34 34170.5 34168.9 16207.1 0.08 0.08 2.11 9 54.17 42723.8 42721.0 20084.3 0.11 0.11 2.13 10 65.01 51234.3 51234.5 23932.7 0.04 0.05 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3680 to 110128 e- Average count rate measured: 788 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.26 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay