*** MIDAS logfile (05SEPpl1.2) *** Th, 31 Aug 2006 14:13:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 31 Aug 2006 14:14:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060831 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.14 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060831 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1701.9 1702.0 862.8 0.11 0.11 1.97 2 4.34 3392.9 3393.1 1709.6 -0.18 -0.17 1.98 3 8.67 6789.9 6790.0 3380.4 -0.07 -0.07 2.01 4 13.00 10181.4 10181.4 5069.3 -0.10 -0.10 2.01 5 19.51 15278.7 15279.1 7547.6 -0.05 -0.05 2.02 6 26.01 20381.1 20380.8 9916.9 -0.00 -0.01 2.06 7 32.51 25486.7 25486.6 12212.3 0.04 0.04 2.09 8 43.34 34000.2 33999.3 16139.6 0.09 0.09 2.11 9 54.17 42504.4 42503.9 20012.2 0.10 0.10 2.12 10 65.01 50982.7 50984.4 23807.3 0.06 0.06 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3658 to 109574 e- Average count rate measured: 784 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.01 e-/ADU RON = 4.25 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay